Bioanalyzer Assays
The RNA 6000 Nano LabChip and the DNA 1000 LabChip are commonly used by the Core. Information for these LabChips is listed here. Information regarding additional LabChips with different sizing and sensitivities are also available directly from Agilent. The Core will train investigators to run the Bioanalyzer themselves.
RNA 6000 Nano
Analysis run time: 30 minutes
Specifications
| |
Total RNA Assay |
mRNA Assay |
| Quantitative range |
25 - 500 ng/ul |
25 - 500 ng/ul |
| Qualitative range |
5 - 500 ng/ul |
25 - 250 ng/ul |
| Maximum sample buffer strength |
100 mM Tris or 125 mM NaCl or 15 mM MgCl2 |
100 mM Tris or 125 mM NaCl or 15 mM MgCl2 |
DNA 1000
Analysis run time: 35 minutes
Specifications
| |
DNA 1000 Assay |
| Sizing range |
25 - 1000 bp |
| Sizing resolution |
±5 bp for 25-100 bp ±5% for 100-1500 bp ±10% for 1500-7500 bp |
| Sizing accuracy |
±10% |
| Quantitation accuracy |
±20% |
| Concentration range |
0.1 - 50 ng/ul |
| Maximum salt concentration per sample |
125 mM for KCl or NaCl 15 nM for MgCl2 |
|